27.8.2012 14:00 setting up logging for ICOS met sensors; 67m desktop sortcut MastMet67m starts logging at 67m Nokeval logger, Module number 7, comport 14, speed 9600bps ch1 temperature Rotronic mV signal ch2 temperature Rotronic external Pt100 ch3 RH Rotronic mV signal ch16 instrument box temperature Pt100 and the data goes to: C:\MastMET\67m_data ; HeLa 28.8.2012 07:52 setting up logging for Nokeval at 125m. Module number 125, comport 18, speed 9600bps The channel setup is following. The Rotronic signals are nonsense, malfunction. ch1 temperature Rotronic mV signal ch2 temperature Rotronic external Pt100 ch3 RH Rotronic mV signal ch4 Uplooking pyranometer mV signal ch5 Downlooking pyranometer mV signal ch16 instrument box temperature Pt100 and the data goes to: C:\MastMET\67m_data; HeLa 04.09.2012 08:30-12:00 installations at 125m box, Rotronic replaced, now VA528*1048 with modified weather shield there, mV and Pt sensors seem to be in wrong order, channel coefficients has to be changed. 3D thies connected to Moxa iport port1 but not tested yet, powered from 24VDC. The Thies cable has been cut shorter, so not any original cable numbers (tags) anymore. Will be documented separately using the numbers and colors on the cable. HeLa 05.09.2012 08:54 the nokeval 125m channels changed, 1 is t_Pt and 2 t_mV, the extra Pt100 sensor and t_mV is the hygroclip millivolt temperature signal. The actual connections will be changed at mast later, 67m and 125m setups will be same. The up and downlooking pyranometers (Middleton solar SK08) are in mastmet125 channels 4,5 now with millivolt output. Need conversion factors to be set -->w/m2; HeLa 13.09.2012 14:54 at 67m nokeval ch1 and ch2 changed according to 125m setup, ch1 is Pt100 ext and ch2 is temperature mV signal. Instrument box grounding connected. HeLa